Enquiry Now
Back to Products
Philips IntelliVue MP5

Philips IntelliVue MP5

₹650000.00

About the Product:

The Philips IntelliVue MP5 is a Compact patient monitor that can fit into a wide range of environments. The IntelliVue MP5 patient monitor has an 8.4 LCD touch screen display. The display screen offers flexible screen configuration. Showing up to four different waveform measurements simultaneously. The Philips IntelliVue MP5 can measure 12 lead ECG. The MP5 can be connected to an IntelliVue MP20 through an MP90 monitor and can be used like a Multi-Measurement Module (MMS). The system has dedicated configurations for anesthesia, critical and cardiac, and neonatal care. Compact, rugged, lightweight monitor with built-in measurements ECG monitoring using any combination of three to 10 electrodes. 12-lead ECG monitoring with five electrodes using the EASI method or with 10 electrodes using the conventional method. Multi-lead arrhythmia and ST segment analysis at the bedside on all available leads. Respironics Mainstream or Sidestream CO2. FAST SpO2 for accurate performance even with low perfusion. Invasive Pressure and Temperature measurement. Choice of either auscultatory or intra-arterial measurements as a measurement reference for Non-Invasive Blood Pressure.

Full Specifications

Height
9.76? (24.8 cm)
Width
10.2? (25.9 cm)
Depth
7.32? (18.6 cm)
Weight
8.8 lbs (4 kg)
Power Consumption
< 40W average; < 65W Peak
Line Voltage
100 to 240 V ~
Current
1.3 to 0.7A
Frequency
50/60 Hz
Type
SVGA
Size
8.4? (21.3 cm)
Resolution
800 x 600
Refresh Rate
60 Hz
Range
0 to 150 mmHg (0 to 20.0 kPa)
Accuracy
after 2 minutes warmup: For values between 0 and 40 mmHg:2.0 mmHg (0.29 kPa) For values from 41 to 70 mmHg: 5% of reading For values from 71 to 100 mmHg: 8% of reading The specifications are valid for standard gas mixtures, balance air, fully hydrated at 35C, Pabs = 760 mmHg, flow rate = 2 l/min
Resolution
Numeric: 1.0 mmHg (0.1 kPa); Wave: 0.1 mmHg (0.01 kPa)
Stability
Short-term drift: 0.8 mmHg over four hours; Long-term drift: Accuracy specification will be maintained over a 120 hour period